Infant stationary incubator SI - 600

Infant stationary incubator SI - 600

Advantages:
  • wall thickness 10 mm - easier and faster cleaning
  • doors opening from the back and from the rear
  • tilting top cover by cleaning
  • alarm signalling visible from long distance
  • alarm priority setting, analyses seriousness of signalled alarms
  • easy patient handling (rotational and withdrawing bed)
  • body temperature also measurable in "Air" mode
  • air curtain with open door

Technical specification

Power voltage 230 V ± 10 %, ~ 50/60
Power input 260 l
- with fixed height 350 VA
- with adjustable height 400 VA
Dimensions
- height without adjusyable leg 1410 mm
- height with adjustable leg 1310 - 1510 mm
- length 950 mm
- width 560 mm
Max. CO2 concentration in patients space 0,4 %

Alarm signals

  • incubator power failure
  • electronic system failure
  • temperature, O2, RH sensor brekadown
  • body/air temperature drop or rise outside adjusted limits
  • temperature rise to emergency level
  • ventilator breakdown
  • heating disconnection
  • O2 concentration drop or rise outside adjusted limits
  • RH drop or rise outside adjusted limits
  • humidifier breakdown

TEMPERATURE Mode

Air temperature mode

20,0 °C - 37,0 °C, v krokoch po 0,1 °C
37,1 °C - 38,0 °C, v krokoch po 0,1 °C

Air temperature setting over 37.0 °C
Automatic adjustment of lower and upper alarm limits

upon special operator's intervention only by pressing the >37 °C key
±1 °C from adjusted required temperature

Body temperature mode

34.0 °C - 37.0 °C, in 0.1 °C steps
37.1 °C - 38.0 °C, in 0.1 °C steps

Warming-up time by 11 °C

30 minutes

 

 

O2 MODE

O2 concentration adjustment

21.5 % - 90 %, in 0.1 % steps

Maximum inside O2 concentration

88 %

O2 regulation accuracy

±1.5 % from adjusted value

Relative humidity at O2 measurement

0 - 99 %

RH MODE

RH concentration adjustment

41 % - 90 %, in 1 % steps

Maximum inside RH

90 %

RH regulation accuracy

± 5 % from adjusted value

Water tank

1.5 Litre sterilized water